Reported Mar 12, 2025 · Ongoing

MiniMed insulin pump, REF: 700G (MMT-1801, MMT-1805, MMT-1850, MMT-1851)

Medtronic MiniMed, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Medtronic MiniMed, Inc. · 18000 Devonshire St · Northridge, CA · 91325-1219

Is this the product you have?

  • MiniMed insulin pump
  • REF
    • 700G (MMT-1801
    • MMT-1805
    • MMT-1850
    • MMT-1851)
Lot codes and dates
  • All Lot Numbers; All UDI numbers
Amount recalled
4,959 unit

This recall covers 12 products

The agency filed one notice per product. You are reading one of them; the other 11 are below.

Where it was sold

Why it was recalled

All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Medtronic MiniMed, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:59 PM UTC