Reported Apr 4, 2018 · Terminated

EasyCAL HbA1c Calibrator; REF 10658 Kit is packaged in a paper box containing 4 vials of 0.25ml of calibrant, one for…

Medica Corporation

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Medica Corporation · 5 Oak Park Dr · Bedford, MA · 01730-1430

Is this the product you have?

  • EasyCAL HbA1c Calibrator; REF 10658 Kit is packaged in a paper box containing 4 vials of 0.25ml of calibrant, one for each calibration level, and an insert sheet.
Lot codes and dates
  • Medica s Lot # 16292

Where it was sold

Why it was recalled

Patient results determined using this calibrator lot might deviate up to 20%. Controls might not indicate false values.

This product could cause temporary or reversible health problems — stop using it.