Reported May 15, 2019 · Terminated
Medica Wash1-Wash solution used prior to Lipase assay on the EasyRA clinical chemistry analyzer REF 10680 - Product…
Medica Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Medica Corporation · 5 Oak Park Dr · Bedford, MA · 01730-1430
Is this the product you have?
- Medica Wash1-Wash solution used prior to Lipase assay on the EasyRA clinical chemistry
- analyzer REF 10680 - Product UsageFor use in washing the probe of the Medica EasyRA Chemistry Analyzer. The wash solution in the wedge is used to prevent interference between certain tests
- Lot codes and dates
Lot #s 18066181331823518317
- Amount recalled
- 87 kits
Where it was sold
- Nationwide
- GA
Why it was recalled
Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer.