Reported May 29, 2019 · Terminated
NeuViz 64 Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body…
Neusoft Medical Systems Co., Ltd.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Neusoft Medical Systems Co., Ltd. · No. 177-1 Cuangxin Road · Shenyang, N/A · N/A
Is this the product you have?
- NeuViz 64 Multi-slice
- CT Scanner System Product UsageThe Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system featuring a continuously rotating x-ray tube and detector array. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles
- Lot codes and dates
37 of 37 products
Products covered by this recall # Product as published 1 N64E120022E 2 N64E130039E 3 N641130051 E 4 N64E130062E 5 N64E130063E 6 N64E130066E 7 1 N641140077E 8 N64E140116E 9 N641140117E 10 N641140118E 11 N641140119E 12 N641140120E 13 N64EN150043E 14 N64E150121E 15 N64E150125E 16 N641150129E 17 N641150131 E 18 N641150136E 19 N641N160066E 20 N641N160067E - Amount recalled
- 37
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Where it was sold
- Nationwide
Why it was recalled
Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, resulting in an incorrect flying focal spot sequence. If the system does not receive error, it would continue to complete the scan with the incorrect flying focal spot sequence, having an effect on the reconstruction of some images, resulting in image loss.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Neusoft Medical Systems Co., Ltd. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.