Reported May 29, 2019 · Terminated
NeuViz 16 Essence Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole…
Neusoft Medical Systems Co., Ltd.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Neusoft Medical Systems Co., Ltd. · No. 177-1 Cuangxin Road · Shenyang, N/A · N/A
Is this the product you have?
- NeuViz 16 Essence Multi-slice
- CT Scanner System Product UsageThe Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system featuring a continuously rotating x-ray tube and detector array. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles
- Lot codes and dates
Series numbers N16E180037EN16E180047E
- Amount recalled
- 2
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Where it was sold
- Nationwide
Why it was recalled
Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, resulting in an incorrect flying focal spot sequence. If the system does not receive error, it would continue to complete the scan with the incorrect flying focal spot sequence, having an effect on the reconstruction of some images, resulting in image loss.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Neusoft Medical Systems Co., Ltd. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.