Reported Mar 31, 2021 · Completed
Biosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (medium) C3, REF: D138502, Rx Only, CE 2797,…
Biosense Webster, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Biosense Webster, Inc. · 15715 Arrow Hwy · Irwindale, CA · 91706-2006
Is this the product you have?
- Biosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (medium) C3
- REFD138502, Rx Only, CE 2797, STERILE EO
- UDI(01)10846835016277 -
- Product Usageprovide accessibility and maneuverability in the cardiac anatomy during cardiac electrophysiological mapping and ablation procedures
- Lot codes and dates
- All Lots
- Amount recalled
- N/A
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Why it was recalled
There is a potential for the hemostatic valve to dislodge while introducing the dilator or device into the sheath. This could result in minor bleeding and/or embolism
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Biosense Webster, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.