Reported Apr 5, 2023 · Ongoing
ARTIS pheno systems with a Siemens Healthineers table or a Trumpf/ MAQUET table-A Fluoroscopic X-Ray system. Material #…
Siemens Medical Solutions USA, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Siemens Medical Solutions USA, Inc · 40 Liberty Blvd · Malvern, PA · 19355-1418
Is this the product you have?
- ARTIS pheno systems with a Siemens Healthineers table or a Trumpf/ MAQUET table-A Fluoroscopic X-Ray system. Material # 10849000.
- Lot codes and dates
- UDI #: 4056869046877
- Serial Numbers
141 of 141 codes
164011164013164018164019164021164022164024164025164026164027164028164033164034164048164049164053164054164055164057164060164061164063164064164066164068164069164071164072164073164074164075164080164083164084164086164094164095164097164098164100164105164113164115164121164124164125164134164138164140164142164152164153164155164157164158164159164160164161164163164168
- Amount recalled
- 141 units US
Where it was sold
- Nationwide
Why it was recalled
In the event of any unintended table movement, the system may not detect the incorrect direction, could lead to the injury of a patient, staff member, operator, or equipment.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Siemens Medical Solutions USA, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.