SOMATOM Definition AS, a family of CT systems intended to produce cross-sectional images of the body by computer…
Siemens Medical Solutions USA, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Siemens Medical Solutions USA, Inc · 40 Liberty Blvd · Malvern, PA · 19355-1418
Is this the product you have?
- SOMATOM Definition AS, a family of CT systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission dta from either the same axial plane taken at different angles or spiral planes taken at different angles.
- Lot codes and dates
- Model NumberSOMATOM Definition AS-8098027 Serial # SOMATOM Definition AS- 95766
- Amount recalled
- 1 CT system
Where it was sold
Why it was recalled
Siemens became aware that the front cover on some Siemens CT system was not properly welded. It is possible for the welded-in threaded bolt to break out of the ground plate of the front cover if the cover is opened during service. This could not happen during clinical operation. Therefore, there is no potential risk for the user or patient and the CT System has no risk for a malfunction during clinical operation
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Siemens Medical Solutions USA, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.