Reported Jul 20, 2022 · Ongoing
Wright EVOLVE TRIAD Plate Cutter, REF 49510120
Wright Medical Technology, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Wright Medical Technology, Inc. · 1023 Cherry Rd · Memphis, TN · 38117-5423
Is this the product you have?
- Wright EVOLVE TRIAD Plate Cutter, REF 49510120
- Lot codes and dates
- Lot codes
9554011168853134930219569362486857
- Lot codes
- Amount recalled
- 231
Where it was sold
- Nationwide
- PR
Why it was recalled
The plate cutter is unable to meet sterilization requirements when it is sterilized in a double-wrapped configuration according to the Option 1 Steam Sterilization parameters detailed in instructions for use
This product could cause temporary or reversible health problems — stop using it.