Reported Apr 20, 2016 · Terminated
Boston Scientific TSX Transseptal Needle, Model Number M004TSX60o; Used to create the primary puncture in the…
Merit Medical Systems, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Merit Medical Systems, Inc. · 1600 West Merit Parkway · South Jordan, UT · 84095
Is this the product you have?
- Boston Scientific TSX Transseptal Needle, Model Number M004TSX60o; Used to create the primary puncture in the interatrial septum when passing an intruder and/or catheter through the septum from the right side of the heart to the left side.
- Lot codes and dates
Lot Q823423Exp 7/31/2018
- Amount recalled
- 25
Why it was recalled
Merit Medical Systems, Inc. is voluntarily conducting a recall of one lot of the TSX" Transseptal Needle due to a discrepancy between the labeling and the actual configuration of the device in the packaging. The curve dimensional values on the labels of this lot incorrectly indicate a 50¿ curve, rather than the correct 86¿ curve
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Merit Medical Systems, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.