Reported Apr 14, 2021 · Terminated

Trial Cutting Guide for Triathlon Total Knee System, Triathlon PKR System, Restoris MCK Knee System, prosthesis, knee,…

Howmedica Osteonics Corp.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Howmedica Osteonics Corp. · 325 Corporate Dr · Mahwah, NJ · 07430-2006

Is this the product you have?

  • Trial Cutting Guide for Triathlon Total Knee System, Triathlon PKR System, Restoris MCK Knee System, prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer
  • Catalog NumberI-K3834CG00 Specialty Triathlon TS Trial Cutting Guide Set I-K3834CG1L Specialty Triathlon TS Trial Cutting Guide Size 1 Left I-K3834CG1R Specialty Triathlon TS Trial Cutting Guide Size 1 Right I-K3834CG2L Specialty Triathlon TS Trial Cutting Guide Size 2 Left I-K3834CG2R Specialty Triathlon TS Trial Cutting Guide Size 2 Right I-K3834CG3L Specialty Triathlon TS Trial Cutting Guide Size 3 Left I-K3834CG3R Specialty Triathlon TS Trial Cutting Guide Size 3 Right I-K3834CG4L Specialty Triathlon TS Trial Cutting Guide Size 4 Left I-K3834CG4R Specialty Triathlon TS Trial Cutting Guide Size 4 Right I-K3834CG5L Specialty Triathlon TS Trial Cutting Guide Size 5 Left I-K3834CG5R Specialty Triathlon TS Trial Cutting Guide Size 5 Right I-K3834CG6L Specialty Triathlon TS Trial Cutting Guide Size 6 Left I-K3834CG6R Specialty Triathlon TS Trial Cutting Guide Size 6 Right I-K3834CG7L Specialty Triathlon TS Trial Cutting Guide Size 7 Left I-K3834CG7R Specialty Triathlon TS Trial Cutting Guide Size 7 Right I-K3834CG8L Specialty Triathlon TS Trial Cutting Guide Size 8 Left I-K3834CG8R Specialty Triathlon TS Trial Cutting Guide Size 8 Right
Lot codes and dates
  • All Lot numbers for each GTINGTIN: 07613327327595
  • GTIN07613327327533
  • GTIN07613327327748
  • GTIN07613327329209
  • GTIN07613327329193
  • GTIN07613327329100
  • GTIN07613327329087
  • GTIN07613327329063
  • GTIN07613327328899
  • GTIN07613327328868
  • GTIN07613327328707
  • GTIN07613327328691
  • GTIN07613327328592
  • GTIN07613327328530
  • GTIN07613327328639
  • GTIN07613327328424
  • GTIN07613327328288
Amount recalled
113 components

Where it was sold

Why it was recalled

The posterior condyle of the Specialty Triathlon TS Trial Cutting Guide(s), sizes 1 through 8, (Left and Right) has the potential to fracture at the augment cut slot. This can occur in the Trial Cutting Guide packaging during transport or during impaction of the instrument in surgery.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Howmedica Osteonics Corp. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated May 10, 2024 · Data as of Sep 19, 7:36 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.