Reported Aug 3, 2022 · Ongoing

CereLink ICP Monitor; Catalog No. 826820, 826820P. Intracranial pressure monitor.

Integra LifeSciences Corp.

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Integra LifeSciences Corp. · 1100 Campus Rd · Princeton, NJ · 08540-6650

Is this the product you have?

    • CereLink ICP Monitor
    • Catalog No. 826820
    • 826820P. Intracranial pressure monitor
Lot codes and dates
  • UDI-DI- 10381780533778; All serial numbers.
Amount recalled
1,210 units

Where it was sold

  • Nationwide
  • GA

Why it was recalled

Stop using this product now

Integra has received complaints associated with ICP readings drifting to -50 mmHg (out-of-range) and may manifest an error message "sensor or extension cable failure!" The firm has found that this is caused by electrical interference from a component of the monitor's circuit board and from the environment.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Integra LifeSciences Corp. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:36 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.