Reported Apr 25, 2018 · Terminated
Ingenuity CT - Model no. 728326, 728325, Computed Tomography X-Ray system Product Usage: Computed Tomography X-ray…
Philips Medical Systems (Cleveland) Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips Medical Systems (Cleveland) Inc · 595 Miner Rd · Cleveland, OH · 44143-2131
Is this the product you have?
Ingenuity CT - Model no. 728326728325Computed
- Tomography X-Ray system Product UsageComputed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories
- Lot codes and dates
- Serial No. :
172 of 172 codes
32045520623000330004300225300293000730022030021130022930024330020930027300413004030024730023330023730023630038300222300227300234300383003730039300483002403004630044300363002153002013002003361213002023002283002483001130021430021730021630022430024230024530023130023030022630023930024130028300423001830019300183002383003033608433305730043
- Serial No. :
- Amount recalled
- 172
This recall covers 14 products
The agency filed one notice per product. You are reading one of them; the other 13 are below.
1 more are in the archive.
Where it was sold
- Nationwide
Why it was recalled
Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Philips Medical Systems (Cleveland) Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.