Reported Jun 19, 2013 · Terminated

ECHELON 60mm Endoscopic Linear Cutter Reloads Black in Flexible Blister pack with Tyvek lid. The ECHELON and ECHELON…

Ethicon Endo-Surgery Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Ethicon Endo-Surgery Inc · 4545 Creek Rd · Cincinnati, OH · 45242-2803

Is this the product you have?

  • ECHELON 60mm Endoscopic Linear Cutter Reloads Black in Flexible Blister pack with Tyvek lid. The ECHELON and ECHELON FLEX families of Endoscopic Linear Cutters (articulating and straight) are sterile, single patient use instruments that simultaneously cut and staple tissue. There are six staggered rows of staples, three on either side of the cut line. The ECHELON 60 instruments have a staple line that is approximately 60 mm long and a cut line that is approximately 57 mm long. The shaft can rotate freely in both directions and an articulation mechanism on articulating instruments enables bending the distal portion of the shaft to facilitate lateral access of the operative site. The instruments are shipped without a reload and must be loaded prior to use. A staple retaining cap on the reload protects the staple leg points during shipping and transportation. The instruments'¿" lock-out feature is designed to prevent a used reload from being refired.
Lot codes and dates
  • The following lot numbers and associated expiration
  • dates are affected by this recall
    • J4AY8Z 5/15/2013
    • J4C00J 6/15/2013
    • J4C87R 7/15/2013
    • J4CC1F 8/15/2013
    • J4CD2X 8/15/2013
    • J4CF3D 9/15/2013
    • J4CJ0E 9/15/2013
    • J4CJ78 9/15/2013
    • J4CK2P 9/15/2013
    • J4CM9X 10/15/2013
    • J4CR2H 10/15/2013
    • J4CT5H 11/15/2013
    • J4CU86 11/15/2013
    • J4CV1W 11/15/2013
    • J4CV5H 11/15/2013
    • K4C149 12/15/2013
    • K4C17W 12/15/2013
    • K4C33U 12/15/2013
    • K4C41C 12/15/2013
    • K4C600 1/15/2014
    • K4C64A 1/15/2014
    • 4C728 1/15/2014
    • K4C82K 1/15/2014
    • K4C88F 1/15/2014
    • K4C94W 1/15/2014
    • K4CC0L 2/15/2014
    • CC8L 2/15/2014
    • K4CD1E 2/15/2014
    • K4CF1P 2/15/2014
    • K4CF5D 2/15/2014
    • K4CG18 3/15/2014
    • K4CG4J 3/15/2014
    • K4CG8P 3/15/2014
    • K4CH44 3/15/2014
    • K4CK0W 3/15/2014
Amount recalled
57,540 Units

Where it was sold

  • Nationwide
  • PR

Why it was recalled

Ethicon Endo-Surgery is initiating a voluntary worldwide recall for ECHELON 60mm Black Reload (ECR60T) due to the potential for incomplete staple line formation from reload damage during the firing sequence.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Ethicon Endo-Surgery Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Feb 11, 2014 · Data as of Sep 19, 4:52 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.