Reported Apr 1, 2020 · Terminated
ADVIA Chemistry Direct Bilirubin_2 assay-SMN: 10316610 (20 mL) & 10341114 (70 mL) - Product Usage: ADVIA DBIL_ 2 is for…
Siemens Healthcare Diagnostics, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Siemens Healthcare Diagnostics, Inc. · 500 Gbc Dr · Newark, DE · 19702-2466
Is this the product you have?
- ADVIA Chemistry Direct Bilirubin_2
- assay-SMN10316610 (20
- mL) & 10341114 (70 mL) - Product UsageADVIA DBIL_ 2 is for in vitro diagnostic use in the quantitative determination of direct bilirubin in human serum and plasma on the ADVIA¿ Chemistry systems. Such measurements are useful in the screening of liver function disorders or in the diagnosis of jaundice. In vitro Diagnostic for the determination of direct bilirubin
- Lot codes and dates
- ADVIA CH DBIL_2 (20 mL) 492068 (01)00630414513089(10)492068(17)20210728 ADVIA CH DBIL_2 (20 mL) 486501 (01)00630414513089(10)486501(17)20210628 ADVIA CH DBIL_2 (20 mL) 478164 (01)00630414513089(10)478164(17)20210328 ADVIA CH DBIL_2 (20 mL) 472667 (01)00630414513089(10)472667(17)20210228 ADVIA CH DBIL_2 (20 mL) 469068 (01)00630414513089(10)469068(17)20201228 ADVIA CH DBIL_2 (20 mL) 462448 (01)00630414513089(10)462448(17)20201028 ADVIA CH DBIL_2 (20 mL) 455476 (01)00630414513089(10)455476(17)20200728 ADVIA CH DBIL_2 (20 mL) 452841 (01)00630414513089(10)452841(17)20200628 ADVIA CH DBIL_2 (20 mL) 447337 (01)00630414513089(10)447337(17)20200528 ADVIA CH DBIL_2 (20 mL) 436518 (01)00630414513089(10)436518(17)20200228 ADVIA CH DBIL_2 (70 mL) 486509 (01)00630414513065(10)486509(17)20210628 ADVIA CH DBIL_2 (70 mL) 478169 (01)00630414513065(10)478169(17)20210328 ADVIA CH DBIL_2 (70 mL) 472672 (01)00630414513065(10)472672(17)20210228 ADVIA CH DBIL_2 (70 mL) 469073 (01)00630414513065(10)469073(17)20201228 ADVIA CH DBIL_2 (70 mL) 462453 (01)00630414513065(10)462453(17)20201028 ADVIA CH DBIL_2 (70 mL) 462459 (01)00630414513065(10)462459(17)20201028 ADVIA CH DBIL_2 (70 mL) 455481 (01)00630414513065(10)455481(17)20200728 ADVIA CH DBIL_2 (70 mL) 452846 (01)00630414513065(10)452846(17)20200628 ADVIA CH DBIL_2 (70 mL) 447351 (01)00630414513065(10)447351(17)20200528 ADVIA CH DBIL_2 (70 mL) 447357 (01)00630414513065(10)447357(17)20200528 ADVIA CH DBIL_2 (70 mL) 458975 (01)00630414513065(10)458975(17)20200528 ADVIA CH DBIL_2 (70 mL) 436523 (01)00630414513065(10)436523(17)20200228
- Amount recalled
- 1901 Us and 37349 OUS
This recall covers 7 products
The agency filed one notice per product. You are reading one of them; the other 6 are below.
Where it was sold
- Nationwide
- GA
Why it was recalled
Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.