Reported May 4, 2016 · Terminated

SOMATOM Force, Computed Tomography x-ray system intended to generate and process cross-sectional images of patients by…

Siemens Medical Solutions USA, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Medical Solutions USA, Inc · 40 Liberty Blvd · Malvern, PA · 19355-1418

Is this the product you have?

  • SOMATOM Force, Computed Tomography x-ray system intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission.
Lot codes and dates
    • SOMATOM Force Material # 10742326
    • serial numbers 75550
    • 75526
    • 75512
    • 75567
    • 75536
    • 75514
    • 75563
    • 75510
    • 75454
    • 75540
    • 75571
    • 75437
    • 75500
    • 75527
    • 75475
    • 75535
    • 75481
    • 75439
    • 75450
    • 75482
    • 75460
    • 75493
    • 75476
    • 75458
    • 75528
    • 75467
    • 75487
    • 75478
    • 75513
    • 75532
    • 75555
    • 75524
    • 75515
    • 75559
    • 75501
    • 75570
    • 75576
    • 75450
    • 75482
    • 75528
    • 75576
Amount recalled
37 CTsystems

Where it was sold

Why it was recalled

The Neonate Head protocol with the pediatric kernel Hp38 could result in artefacts and possibly lead to a misdiagnosis (i.e. either non-existing blood or liquid is mimicked in the images or actual existing blood or liquid is not depicted as expected). There is also a risk of a potential misdiagnosis in using this protocol for surgery planning. Investigations revealed an incorrect parameterization of the reconstruction algorithm applied for the pediatric head kernel as cause of the problem.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Medical Solutions USA, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 17, 2017 · Data as of Sep 19, 4:44 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.