Reported May 11, 2016 · Terminated
MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking…
Siemens Medical Solutions USA, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Siemens Medical Solutions USA, Inc · 40 Liberty Blvd · Malvern, PA · 19355-1418
Is this the product you have?
- MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.
- Lot codes and dates
- Model # 10432914
- and 10432915 Serial Numbers
4247242489424774616946178461754247842485
- Amount recalled
- 8 systems
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Siemens Medical Solutions USA, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.