Reported Apr 15, 2020 · Terminated
Component for Laser Scanning Microscopes (LSM 700, LSM 800 and LSM 900)
Zeiss, Carl Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Zeiss, Carl Inc · 1 Zeiss Dr · Thornwood, NY · 10594-1939
Is this the product you have?
- Component for Laser Scanning Microscopes (LSM 700, LSM 800 and LSM 900)
- Lot codes and dates
(LSM 700LSM 800 and LSM 900)
- Amount recalled
- US: 600; Global: 1600
Where it was sold
- Nationwide
Why it was recalled
When the user tilts the transmitted light arm of the microscope to the end of its axis range, laser safety can be bypassed and result in emission of class lIlb laser radiation. This issue arises when a ZEISS inverted stage stand (Axio Observer or Axio Vert) is used in as a component of the LSM 700, 800 and 900 system