Reported Jul 25, 2018 · Terminated

Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope

Zeiss, Carl Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Zeiss, Carl Inc · 1 Zeiss Dr · Thornwood, NY · 10594-1939

Is this the product you have?

  • Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope
Lot codes and dates
  • 432332-9020-000
Amount recalled
8

Where it was sold

  • Nationwide

Why it was recalled

Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Zeiss, Carl Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jul 26, 2024 · Data as of Sep 19, 4:47 PM UTC