Reported Sep 7, 2022 · Ongoing

Artis icono biplane- Model No. 11327600 Artis icono floor- Model No. 11327700

Siemens Medical Solutions USA, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Medical Solutions USA, Inc · 40 Liberty Blvd · Malvern, PA · 19355-1418

Is this the product you have?

  • Artis icono biplane- Model No. 11327600 Artis icono floor- Model No. 11327700
Lot codes and dates
  • Artis icono biplane- Model No. 11327600
  • UDI-DI04056869063317 Serial Numbers

    77 of 77 codes

    • 180062
    • 180054
    • 180019
    • 180421
    • 180444
    • 180034
    • 180072
    • 180446
    • 180367
    • 180438
    • 180382
    • 180060
    • 180428
    • 180369
    • 180370
    • 180371
    • 180372
    • 180080
    • 180301
    • 180305
    • 180067
    • 180399
    • 180442
    • 180040
    • 180379
    • 180386
    • 180360
    • 180333
    • 180352
    • 180407
    • 180433
    • 180412
    • 180400
    • 180397
    • 180432
    • 180316
    • 180447
    • 180363
    • 180429
    • 180084
    • 180418
    • 180334
    • 180018
    • 180376
    • 180354
    • 180046
    • 180434
    • 180357
    • 180358
    • 180361
    • 180362
    • 180081
    • 180423
    • 180411
    • 180414
    • 180381
    • 180021
    • 180443
    • 180309
    • 180356
    ***Updated 9/23/22*** 180466 180465 180464 180047 180463 ***Updated 1/9/23***
    • 180470
    • 180534
    • 180468
    • 180472
    • 180467
    • 180469
    • 180066
    • 180082
    Artis icono floor- Model No. 11327700
  • UDI-DI04056869149325 Serial Numbers
    • 170319
    • 170391
    • 170392
    • 170397
    • 170026
    • 170388
    • 170343
    • 170367
    • 170368
    • 170360
    • 170349
    • 170361
    • 170383
    • 170322
    • 170385
    • 170386
    • 170025
    • 170384
    • 170346
    • 170051
    • 170358
    • 170389
    • 170357
    • 170365
    • 170387
    • 170390
    • 170393
    • 170394
    • 170378
    • 170396
    • 170053
    • 170364
    ***Updated 1/9/23*** 170402
Amount recalled
1,389 (590 US; 799 OUS) (Total)

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

  • Nationwide

Why it was recalled

Siemens has become aware of a potential hardware issue with Artis Q, Artis Q.zen, Artis pheno and Artis icono systems with a specific lot of X-ray tubes. In rare cases, if the error detection mechanism fails, it will not be possible to release X-ray any longer until system shutdown.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Medical Solutions USA, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:54 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.