Reported May 18, 2016 · Terminated
Advia Chemistry Assays - Cholesterol_2 concentrated and Cholesterol, LDL Cholesterol
Siemens Healthcare Diagnostics, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Siemens Healthcare Diagnostics, Inc. · 500 Gbc Dr Ms 514 · Newark, DE · 19702-2466
Is this the product you have?
- Advia Chemistry Assays - Cholesterol_2 concentrated and Cholesterol, LDL Cholesterol
- Lot codes and dates
- Cholesterol_2 (CHOL2): 74719 Cholesterol_concentrated (CHOL_c): 74718
- Amount recalled
- 137422 units total
This recall covers 20 products
The agency filed one notice per product. You are reading one of them; the other 19 are below.
7 more are in the archive.
Where it was sold
Why it was recalled
Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depressed results may occur on samples drawn from patients receiving N-Acetylcysteine (NAC) or Metamizole. NAC is the accepted antidote for acetaminophen toxicity and is justified in patients at signific
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.