Reported Apr 15, 2020 · Terminated
PRIMUS HI #04504200 with Digital LINAC Systems Control Console Software Versions 6, 7, 9, 13
Siemens Medical Solutions USA, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Siemens Medical Solutions USA, Inc · 40 Liberty Blvd · Malvern, PA · 19355-1418
Is this the product you have?
- PRIMUS HI #04504200 with Digital LINAC Systems Control Console Software Versions 6, 7, 9, 13
- Lot codes and dates
Serial Numbers 302530493241326733053422347935153519354735653602373137513832366739243949403670-418551533361397936573679
- Amount recalled
- 230,001
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
Why it was recalled
Potential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy.
This product could cause temporary or reversible health problems — stop using it.