Reported May 16, 2018 · Terminated

ADVIA Centaur¿ Cyclosporine

Siemens Healthcare Diagnostics, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Healthcare Diagnostics, Inc. · 511 Benedict Ave · Tarrytown, NY · 10591-5005

Is this the product you have?

  • ADVIA Centaur¿ Cyclosporine
Lot codes and dates
  • ADVIA Centaur¿ Cyclosporine; Cyclosporine
  • SMN 10335448 Lot codesADVIA Centaur CSA Lot #
    • 38182025
    • 64432025
    • 98554025
    • 58299030
    • 25547030
    • 30153031
    • 77934031
    • 97170031
Amount recalled
N/A

This recall covers 44 products

The agency filed one notice per product. You are reading one of them; the other 43 are below.

31 more are in the archive.

Where it was sold

  • Nationwide

Why it was recalled

The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Feb 10, 2023 · Data as of Sep 19, 6:15 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.