Reported Jun 10, 2015 · Terminated
Brilliance iCT Computed Tomography X-ray system
Philips Medical Systems (Cleveland) Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips Medical Systems (Cleveland) Inc · 595 Miner Rd · Cleveland, OH · 44143-2131
Is this the product you have?
- Brilliance iCT Computed Tomography X-ray system
- Lot codes and dates
Model# 728306Serial #'s: 85019 & 85023
- Amount recalled
- 2
Why it was recalled
A calibrated torque wrench malfunctioned at the supplier leading to possible incorrect torque of x-ray tube mounting fasteners.