Reported Apr 12, 2017 · Terminated

ADVIA Chemistry Enzyme 2 Calibrator, multi-analyte mixture Product Usage: Assays included in Enzyme 2 Calibrator: AST,…

Siemens Healthcare Diagnostics, Inc.

Low riskThis product is unlikely to cause harm, but it breaks a safety or labelling rule.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Healthcare Diagnostics, Inc. · 500 Gbc Dr Ms 514 · Newark, DE · 19702-2466

Is this the product you have?

  • ADVIA Chemistry Enzyme 2 Calibrator
  • multi-analyte mixture Product UsageAssays
  • included in Enzyme 2 CalibratorAST, AST_c ALT, ALT_c, ALTP5P, ALTP_c, ASTP5P and ASTP_c
Lot codes and dates
  • Lots 6JD079 and 6MD042
Amount recalled
314 calibrators

Where it was sold

  • Nationwide

Why it was recalled

Siemens Healthcare Diagnostics has confirmed a positive bias of up to 41% in quality control and patient results for Aspartate Aminotransferase (AST) and Aspartate Aminotransferase concentrated (AST_c) assays following calibration with Advia Chemistry Enzyme 2 Calibrator lots 6JD079 and 6MD042.

This product is unlikely to cause harm, but it breaks a safety or labelling rule.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated May 9, 2018 · Data as of Sep 19, 4:48 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.