Reported May 16, 2018 · Terminated

Dimension Lipase Flex¿ Reagent Cartridge - ( Lipase-esterase, enzymatic, photometric, lipase)

Siemens Healthcare Diagnostics, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Healthcare Diagnostics, Inc. · 500 Gbc Dr Ms 514 · Newark, DE · 19702-2466

Is this the product you have?

  • Dimension Lipase Flex¿ Reagent Cartridge - ( Lipase-esterase, enzymatic, photometric, lipase)
Lot codes and dates
  • Catalog DF56 Lot FB8109- UDI# (01)00842768025532(10)FB8109(17)20180419
Amount recalled
N/A

This recall covers 44 products

The agency filed one notice per product. You are reading one of them; the other 43 are below.

31 more are in the archive.

Why it was recalled

Calibration failure - confirmed a positive bias with LIPL lot FB8109 when compared to other Dimension LIPL lots. Lot FB8109 is not meeting Siemens internal specifications for lot-to-lot limits.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Dec 3, 2020 · Data as of Sep 19, 4:49 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.