Reported Jun 4, 2025 · Ongoing

ADVIA Centaur HCY Calibrator (2 Pack). Siemens Material Number: 10310376.

Siemens Healthcare Diagnostics, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Healthcare Diagnostics, Inc. · 333 Coney St · East Walpole, MA · 02032-1516

Is this the product you have?

  • ADVIA Centaur HCY Calibrator (2 Pack)
  • Siemens Material Number10310376
Lot codes and dates
  • Siemens Material Number10310376
  • UDI-DI 00630414468198 Kit lot numbers86237A74, 44207A75. ***Updated 5/8/26*** Additional lots 25674A77 52363A79
Amount recalled
1,072 packs (2,144 units) - 156 packs US, 916 packs OUS

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:18 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.