Reported Jul 17, 2019 · Terminated

Giraffe Infant Radiant Warmer Infant radiant warmers provide infrared heat in a controlled manner to neonates who are…

GE Healthcare, LLC

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: GE Healthcare, LLC · 3000 N Grandview Blvd · Waukesha, WI · 53188-1615

Is this the product you have?

  • Giraffe Infant Radiant Warmer Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment.
Lot codes and dates

631 of 631 products

Products covered by this recall
#Product as published
1Model Number, (Serial Number, UDI Number): M1118179-529707 (GBWW60444, 010084068210392321GBWW6044411170620)
2M1118179-529707 (GBWW60446, 010084068210392321GBWW6044611170620)
3M1118179-024201 (HDJM50619, Not Applicable)
4M1118179-024201 (HDJM50620, Not Applicable)
5M1118179-025977 (HDJN50298, Not Applicable)
6M1118179-025977 (HDJN50299, Not Applicable)
7M1118179-029089 (HDJP50554, Not Applicable)
8M1118179-029089 (HDJP50555, Not Applicable)
9M1118179-038813 (HDJR54214, Not Applicable)
10M1118179-038813 (HDJR54215, Not Applicable)
11M1118179-038813 (HDJR54216, Not Applicable)
12M1118179-024011 (HDJS50551, Not Applicable)
13M1118179-024011 (HDJS50552, Not Applicable)
14M1118179-024011 (HDJS50553, Not Applicable)
15M1118179-024011 (HDJS50554, Not Applicable)
16M1118179-024011 (HDJS50555, Not Applicable)
17M1118179-150421 (HDJU61285, Not Applicable)
18M1118179-150421 (HDJU61286, Not Applicable)
19M1118179-150421 (HDJU61287, Not Applicable)
20M1118179-150421 (HDJS52930, Not Applicable)
Amount recalled
9,094 devices

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

Stop using this product now

The bedside panels and latch areas on the Giraffe Bedded Warmer and Panda iRes Warmer can crack or break if the unit is moved using the bedside panels instead of the front handle or the maneuvering handle.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new GE Healthcare, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated May 21, 2024 · Data as of Sep 19, 6:53 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.