Reported May 13, 2020 · Terminated

Visualase Cooled Laser Applicator System Tubing Set kits Product numbers: VCLAS 9735559, 3mm tip/1.65mm catheter VCLAS…

Medtronic Navigation, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Medtronic Navigation, Inc. · 826 Coal Creek Cir · Louisville, CO · 80027-9710

Is this the product you have?

  • Visualase Cooled Laser Applicator
  • System Tubing Set kits Product numbers
    • VCLAS 9735559
    • 3mm tip/1.65mm catheter VCLAS 9735560
    • 10mm tip/1.65mm catheter VCLAS 9735561
    • 15mm tip/1.85mm catheter
Lot codes and dates
  • All kits with lot numbers from 0211041602 through 0217695790
Amount recalled
4654 units

Where it was sold

Why it was recalled

The firm received complaints on the saline tubing of Visualase Cooled Laser Applicator System (VCLAS) kit, including incorrect Luer connector assembly, incorrect drip-chamber assembly, and/or excessive adhesive causing tubing occlusion, all of which could result in disruption or prevention of saline flow.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Medtronic Navigation, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Oct 16, 2024 · Data as of Sep 19, 5:31 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.