Reported Jun 30, 2021 · Terminated
Access SARS CoV-2 lgG II Reagent, REF: C69057, IVD CE UDI: (01)15099590742744
Beckman Coulter Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Beckman Coulter Inc. · 250 S Kraemer Blvd · Brea, CA · 92821-6232
Is this the product you have?
- Access SARS CoV-2 lgG II Reagent
- REFC69057
- IVD CE UDI(01)15099590742744
- Lot codes and dates
- All Lots
- Amount recalled
- 13,337 units
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
SARS-CoV-2 lgG II numerical results may potentially be multiplied by a factor of 1000 on systems running with assay protocol file (APF) and access assay file (AAF) versions resulting in falsely elevated numerical values.