Reported Jun 15, 2016 · Terminated

Triglyceride, Catalog No. OSR6x118 Product Usage:Triglyceride OSR6x118 is intended as an Enzymatic colour test for the…

Beckman Coulter Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Beckman Coulter Inc. · 250 S Kraemer Blvd · Brea, CA · 92821-6232

Is this the product you have?

  • Triglyceride, Catalog No
  • OSR6x118 Product UsageTriglyceride OSR6x118 is intended as an Enzymatic colour test for the quantitative determination of triglyceride in human serum and plasma on Beckman Coulter AU analyzers
Lot codes and dates
  • All lots
Amount recalled
79,009 units total (21,215 units in US)

This recall covers 5 products

The agency filed one notice per product. You are reading one of them; the other 4 are below.

Where it was sold

  • Nationwide
  • PR

Why it was recalled

Beckman Coulter has identified that Dipyrone (Metamizole) may cause erroneous, false low results in the following assays: Cholesterol OSR6X16, Triglyceride OSR6X118, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Beckman Coulter Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jun 30, 2016 · Data as of Sep 19, 6:46 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.