Reported Jun 6, 2018 · Terminated

Syva¿ EMIT¿ 2000 Theophylline Product Usage: The Syva¿ Emit¿ 2000 Theophylline Assay is a homogeneous enzyme…

Siemens Healthcare Diagnostics, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Healthcare Diagnostics, Inc. · 500 Gbc Dr Ms 514 · Newark, DE · 19702-2466

Is this the product you have?

  • Syva¿ EMIT¿ 2000
  • Theophylline Product UsageThe Syva¿ Emit¿ 2000 Theophylline Assay is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of theophylline in human serum or plasma
Lot codes and dates
  • Catalog # 4P019UL /SMN#10445324 kit lot J1 (UDI# 00842768001024J120190414)
Amount recalled
N/A

Where it was sold

  • Nationwide

Why it was recalled

Lot J1 of the Syva¿ EMIT¿ 2000 Theophylline Assay, when run on a Beckman AU Clinical Chemistry System, may exhibit increased imprecision for commercially available controls and for patient samples.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 19, 2019 · Data as of Sep 19, 5:25 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.