Reported Jul 18, 2012 · Terminated
Drager PT 4000 Phototherapy System
Draeger Medical Systems, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Draeger Medical Systems, Inc. · 3135 Quarry Rd · Telford, PA · 18969-1042
Is this the product you have?
- Drager PT 4000 Phototherapy System
- Lot codes and dates
- Catalog number 2M21700 and serial numbers ASCF-0037 TO ASCF-0061; ASCJ-0037 TO ASCJ-0152; ASCL-0001 TO ASCL-0025; ASCM-0001 TO ASCM-0025; and ASDA-002, 0005, 0006, 0007, 0008, 0009, 0012, 0019, 0025.
- Amount recalled
- 65
Where it was sold
- Nationwide
- PR
Why it was recalled
A routine internal quality check found that the Drager PT 4000 Phototherapy System is fitted with a 16 Amp rated fuse instead of the specified 1.6 Amp rated fuse. This may mean that the internal fuse would not trip promptly in the event of an internal short-circuit. The possible consequence is the wall outlet circuit breaker would trip first on the device may severely overheat. There have not b
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Draeger Medical Systems, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.