Reported Jul 18, 2012 · Terminated

GE Advantage CT/PET Fusion for Advantage Windows Workstation with Volume View CT Colonography/Navigator 2 software.…

GE Healthcare, LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: GE Healthcare, LLC · 3000 N Grandview Blvd · Waukesha, WI · 53188-1615

Is this the product you have?

  • GE Advantage CT/PET Fusion for Advantage Windows Workstation with Volume View CT Colonography/Navigator 2 software. (Version 4.0, 4.1,or 4.2) The Advantage CT/PET Fusion clinical application provides easy comparison of three dimensional (3D) anatomical images from Computed Tomography (CT) and functional images from Positron Emission Tomography (PET).
Lot codes and dates
  • 70 of 70 codes

    • 261AWUSV35007H7
    • 10026AW
    • 514934BAR14
    • 26329AW2
    • ACBCTAW7
    • 956632AW42
    • 321799AW1
    • 617355AWPC1
    • 617355AW3
    • 361595SKW1
    • 617754MR5AW
    • 3454AWW
    • 3526
    • 941747AWW
    • 212639XL5
    • 239436AW2
    • 773792AW1
    • 803366AW1
    • 813642AW42
    • 973972AWW3
    • 973972AWW5
    • 210292AW41C
    • 082445030001AW
    • 082445010004AW
    • 082421010160AW
    • 082421120176AW
    • 082406200003AW
    • 082421150050AW
    • 082416130012AW
    • 082445150008AW
    • 082421100265AW
    • 082421050047AW
    • 083049822101584
    • 083055602101433
    • 083055065231713
    • 083049842101695
    • PT0004
    • PT0023
    • 0847490042
    • 0853550153
    • 0853550221
    • AH4511AW01
    • 030070OT02
    • EG1277AW01
    • M4017059
    • PC0125AW02
    • PC0815AW03
    • PC0362AW04
    • IE1005AW02
    • A5248812
    • A5158021
    • A5226504
    • KE1000AW01
    • RU2441AW02
    • SA1054AW01
    • NP083626
    • MP825712
    • FPG72503
    • 0853550150
    • TN1003AW02
Amount recalled
70

Where it was sold

Why it was recalled

GE Healthcare has become aware of a potential safety issue due to incorrect alignment of fused CT and PET images associated with the CT/PET fusion capabilities of your Volume Viewer and CT/PET Fusion application. Volume Viewer allows two studies (one PET and one CT) that are registered in the same spatial domain to be viewed simultaneously in fused mode. The two images should be aligned to th

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new GE Healthcare, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Dec 6, 2012 · Data as of Sep 19, 6:13 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.