Reported Jul 9, 2014 · Terminated

MIS Rod 2, a component of the NewPort System. Model No: 60-0545 (45mm); 60-0550 (50mm); 60-0555 (55mm); 60-0560 (60mm);…

Integra LifeSciences Corp.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Integra LifeSciences Corp. · 311 Enterprise Dr · Plainsboro, NJ · 08536-3344

Is this the product you have?

  • MIS Rod 2, a component of the NewPort System
  • Model No60-0545 (45mm); 60-0550 (50mm); 60-0555 (55mm); 60-0560 (60mm); 60-0575 (75mm); 60-0580 (80mm). The NewPort MIS Rod 2 is a 5.5 diameter titanium spinal rod used to immobilize adjacent NewPort pedicle screws
Lot codes and dates
    • Lot No. Z21781A (45mm NewPort MIS Rod
    • 2) Z21956A (50mm NewPort MIS Rod
    • 2) Z21782A (55mm NewPort MIS Rod
    • 2) Z21957A (60mm NewPort MIS Rod
    • 2) Z21784A (75mm NewPort MIS Rod
    • 2) Z21958A (80mm NewPort MIS Rod
    • 2)
Amount recalled
94 units

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

Why it was recalled

Integra Lifesciences is recalling NewPort System MIS Rod, 2 because it may be up to 30 degrees offset from specification requirements causing misalignment with the inserter. This could potentially result in the lordosis of the rod not aligning with the handle when the inserter is used in accordance with the current NewPort MIS System Surgical Technique.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Integra LifeSciences Corp. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Nov 6, 2014 · Data as of Sep 19, 4:50 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.