Reported Jun 26, 2024 · Ongoing
Ventilator HAMILTON-C6, PN: 160021
Hamilton Medical AG
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Hamilton Medical AG · Via Crusch 8 · Bonaduz, N/A · N/A
Is this the product you have?
- Ventilator HAMILTON-C6
- PN160021
- Lot codes and dates
- UDI07630002808590
- Software Versions
SW v1.1.4SW v1.1.5SW v1.1.6
- Amount recalled
- 68
Where it was sold
Why it was recalled
Stop using this product now
Ventilator may enter sensor fail mode, ventilation may not re-initiate, after patient is reconnected, which may cause hypoxia, if this sequence occurs: 1)User presses O2 enrichment, 2)User disconnects endotracheal tube for open suctioning, 3)Sensor error initiated, sensor fail mode occurs, 4)Patient is re-connected with sensor fail mode active, 5) Ventilation is not re-initiated by the ventilator.
Using this product could cause serious injury, illness or death — stop using it now.