Reported Aug 27, 2025 · Ongoing

Hamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to…

Hamilton Medical AG

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Hamilton Medical AG · Parc Industrial Vial 10 · Domat/Ems, N/A · N/A

Is this the product you have?

  • Hamilton C6 REF 160021 (ESM
  • Board PN 10144626 UDI-DI code76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive pressure ventilatory support to adults and pediatrics and optionally infants and neonates
Lot codes and dates
  • Model/Catalog Number160021
  • ESM board Part Number10144626
  • GTIN07630002808590
  • UDI-DI76300028PN160021ZW
  • Affected SN range13636 till 13950 and 13966 till
    • 14753
    • 13797
    • 13801
    • 13806
    • 13855
    • 13865
    • 13981
    • 13988
    • 13989
    • 14012
    • 14015
    • 14019
    • 14034
    • 14035
    • 14042
    • 14052
    • 14054
    • 14102
    • 14109
    • 14112
    • 14113
    • 14114
    • 14115
    • 14116
    • 14119
    • 14120
    • 14121
    • 14122
    • 14123
    • 14124
    • 14126
    • 14127
    • 14128
    • 14163
    • 14168
    • 14169
    • 14170
    • 14172
    • 14173
    • 14175
    • 14177
    • 14187
    • 14188
    • 14189
    • 14195
    • 14197
Amount recalled
1040 systems

Where it was sold

  • Nationwide

Why it was recalled

Stop using this product now

A mechanical defect on a printed circuit board (PCB) as a result of PCB separating process. This defect might cause the ventilator to either fail at startup (no patient involvement) or during the use of the device enter ambient state, resulting in an interruption of ventilation.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Hamilton Medical AG recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 3:22 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.