Reported Jul 2, 2025 · Ongoing

REMISOL Advance UDI-DI code: 13700962601874 All Software Versions: B92487; B92488; C24317; C28652; C37500; D04164;…

Beckman Coulter Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Beckman Coulter Inc. · 250 S Kraemer Blvd · Brea, CA · 92821-6232

Is this the product you have?

14 of 14 products

Products covered by this recall
#BarcodesProduct as published
113700962601874REMISOL Advance UDI-DI code: 13700962601874 All Software Versions: B92487
2B92488
3C24317
4C28652
5C37500
6D04164
7C44703
8C57017
9C69412
10C69413
11C73942
12C73941
13C88470
14C88471 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections (Beckman Coulter Instruments, automations, LIS&). It can also be connected to a host system. It provides data management capabilities such as automatic results validation, delta checking, reflex testing, quality control, results editing, archiving and restoration of patient results.
Amount recalled
1,805 systems

Where it was sold

Why it was recalled

Due to a software issue, the software may incorrectly process the date of birth (DoB) received from the Laboratory Information System (LIS) and when the DoB is set to 01/01/1900 or left empty, REMISOL Advance may incorrectly identify a patient as a newborn, affecting the execution of validation rules based on incorrect reference ranges. The test results are not changed by this software defect and are accurately displayed by REMISOL Advance; however, the evaluation flagging of the results may be affected. This issue may lead to wrong reference range calculations and can result in erroneous flagged results being auto validated and uploaded to LIS.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Beckman Coulter Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 7:41 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.