Reported Jul 31, 2019 · Terminated

Artis zee floor MN, Model Number 10094142

Siemens Medical Solutions USA, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Medical Solutions USA, Inc · 40 Liberty Blvd · Malvern, PA · 19355-1418

Is this the product you have?

  • Artis zee floor MN, Model Number 10094142
Lot codes and dates
  • Serial Numbers
    • 140212
    • 140115
    • 140101
    • 140205
    • 140213
    • 140204
    • 140209
    • 140219
    • 140224
    • 140218
    • 140221
    • 140220
    • 140206
    • 140223
    • 140105
    • 140226
    • 140108
    • 140102
    • 140109
    • 140100
    • 140202
    • 140214
    • 140114
Amount recalled
23

This recall covers 23 products

The agency filed one notice per product. You are reading one of them; the other 22 are below.

10 more are in the archive.

Where it was sold

  • Nationwide

Why it was recalled

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Medical Solutions USA, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Sep 18, 2020 · Data as of Sep 19, 5:35 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.