Reported Jul 31, 2019 · Terminated

Artis zeego, Model Number 10280959

Siemens Medical Solutions USA, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Medical Solutions USA, Inc · 40 Liberty Blvd · Malvern, PA · 19355-1418

Is this the product you have?

  • Artis zeego, Model Number 10280959
Lot codes and dates
  • Serial Numbers

    117 of 117 codes

    • 160395
    • 160484
    • 160502
    • 160351
    • 160392
    • 160449
    • 160463
    • 160313
    • 160411
    • 160406
    • 160106
    • 160134
    • 160136
    • 160347
    • 160312
    • 160488
    • 160448
    • 160427
    • 160455
    • 160486
    • 160435
    • 160451
    • 160346
    • 160384
    • 160100
    • 160493
    • 160391
    • 160321
    • 160376
    • 160447
    • 160471
    • 160481
    • 160401
    • 160119
    • 160403
    • 160404
    • 160501
    • 160438
    • 160388
    • 160378
    • 160419
    • 160114
    • 160450
    • 160429
    • 160149
    • 160137
    • 160152
    • 160153
    • 160154
    • 160138
    • 160446
    • 160368
    • 160394
    • 160412
    • 160387
    • 160432
    • 160147
    • 160433
    • 160483
    • 160335
Amount recalled
117

This recall covers 23 products

The agency filed one notice per product. You are reading one of them; the other 22 are below.

10 more are in the archive.

Where it was sold

  • Nationwide

Why it was recalled

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Medical Solutions USA, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Sep 18, 2020 · Data as of Sep 19, 6:52 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.