Reported Jun 19, 2024 · Ongoing
MyoSPECT System, Model H3912AA and H3912B, and MyoSPECT ES System, Model H3912AB and H3912BD, dedicated cardiac SPECT…
Ge Medical Systems Israel, Functional Imaging
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING · 4, Hayozma St · Tirat Carmel, N/A · N/A
Is this the product you have?
- MyoSPECT System, Model H3912AA and H3912B, and MyoSPECT ES System, Model H3912AB and H3912BD, dedicated cardiac SPECT (single photon emission computerized tomography system).
- Lot codes and dates
- MyoSPECT
- manufactured on or before 2024-02-13Model H3912AA, GTIN 00195278421586 - Serial numbers NGCC80094, NGCB80033, NGCB80044, NGCB80045, NGCA80017, NGCB80039, NGCA80006, NGCA80011, NGCA80003, NGCA80012, NGCB80052, NGCB80060, NGCC80074, NGCC80081, NGCB80038, NGCD80107, NGCB80070, NGCC80093, NGCD80106, NGCB80026, NGCB80050, NGCB80058, NGCC80084, NGCB80027, NGCB80072, NGCB80032, NGCB80031, NGCC80080, NGCB80067, NGCA80021, NGCA80010, NGCA80015, NGCB80043, NGCC80090, NGCC80091, NGCD80104, NGCB80028, NGCB80022, NGCB80023, NGCB80048, NGCB80035, NGCD80103, NGCC80100, NGCC80083, NGCB80042, NGCC80075, NGCB80068, NGCB80040, NGCB80051, NGCB80064, NGCC80078, NGCC80079, NGCC80087, NGCC80092, NGCD80105, NGCD80108, NGCB80059, NGCA80009, NGCC80086, NGCC80095, NGCA80020, NGCC80088, NGCB80055, NGCB80029, NGCB80041, NGCC80076, NGCA80016, NGCB80036, NGCC80098, NGCB80037, NGCB80046, NGCA80007, NGCC80082, NGCB80061, NGCA80002, NGCC80102, NGCB80057, NGCC80089, NGCB80030, NGCC80097, NGCA80005, NGCB80049, NGCC80101, NGCB80071, NGCA80001, NGCB80066, NGCB80065, NGCC80085, NGCC80077, NGCB80069, NGCA80014, MYGA10001, NGCC80096, NGCB80024, NGCB80047, NGCC80099, NGCB80062, NGCB80034, NGCA80008, NGCA80013, NGCB80025, NGCA80004, NGCA80019, NGCB80056, NGCB80063, NGCB80054, and NGCB80053. Model H3912BC, GTIN 00195278488619 - Serial number MYOC93001. MyoSPECT
- ES manufactured on or before 2024-02-15Model H3912AB, GTIN 00195278421579 - Serial numbers NGEB81002, NGED81025, NGEB81007, NGEB81006, NGEC81019, NGEB81012, NGEB81008, NGEC81014, NGEB81005, NGEC81017, NGEC81022, NGEC81021, NGEC81020, NGEB81011, NGEB81010, NGEC81023, NGEC81024, NGEC81013, NGEB81009, NGEB81003, NGEB81004, NGEC81018, NGEC81015, and NGEC81016. Model H3912BD, GTIN 00195278488626 - Serial number MYEB94001
- Amount recalled
- 133 devices
Where it was sold
Why it was recalled
There is a potential issue regarding the service handles for the internal lead (Pb) covers in that the handles could loosen or release which could cause a cover to drop potentially resulting in injury to service personnel due to the weight of the cover.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Ge Medical Systems Israel, Functional Imaging recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.