Reported Jan 22, 2025 · Ongoing

GE Healthcare Infinia I, Model/Catalog Numbers: 1) 5178728, 2) H3000WE, 3) H3000WM, 4) H3000YG; System, Tomography,…

Ge Medical Systems Israel, Functional Imaging

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING · 4, Hayozma St · Tirat Carmel · N/A

Is this the product you have?

  • GE Healthcare Infinia I
  • Model/Catalog Numbers1) 5178728, 2) H3000WE, 3) H3000WM, 4) H3000YG; System, Tomography, Computed, Emission
Lot codes and dates

35 of 35 products

Products covered by this recall
#BarcodesProduct as published
1UDI/DI N/A, System ID YN1200, Serial Number: 16179
2System ID YN1650, Serial Number: 17339. 2) H3000WE: UDI/DI N/A, System ID 82426060010, Serial Number Not Available
3System ID 82426080021, Serial Number Not Available
4System ID 82426090035, Serial Number 18605
5System ID 82426090037, Serial Number 18604
6System ID 82426100057, Serial Number 19191
7System ID 82426150036, Serial Number 18624
8System ID 82426150038, Serial Number Not Available
9System ID 82426180015, Serial Number 18611
10082426010021System ID 082426010021, Serial Number 18644
11082426030032System ID 082426030032, Serial Number 18627
12082426050012System ID 082426050012, Serial Number 18629
13082426070041System ID 082426070041, Serial Number 18639
14082426090044System ID 082426090044, Serial Number 18652
15082426100067System ID 082426100067, Serial Number 18648
16082426100069System ID 082426100069, Serial Number 18641
17082426120031System ID 082426120031, Serial Number 18656
18082426120033System ID 082426120033, Serial Number 18650
19082426140029System ID 082426140029, Serial Number 18628
20082426140030System ID 082426140030, Serial Number 18647
Amount recalled
36 units

This recall covers 13 products

The agency filed one notice per product. You are reading one of them; the other 12 are below.

Why it was recalled

GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Ge Medical Systems Israel, Functional Imaging recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:54 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.