Reported Jul 12, 2023 · Ongoing
BETTA LINK LG KNOTLESS IMPLANT KIT-Intended for use in soft tissue to bone fixation in the repair of the natural…
T.a.g. Medical Products Corporation, Ltd.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: T.A.G. MEDICAL PRODUCTS CORPORATION, LTD. · Kibbutz Gatton · GAATON, N/A · N/A
Is this the product you have?
- BETTA LINK LG KNOTLESS IMPLANT KIT-Intended for use in soft tissue to bone fixation in the repair of the natural ligament or tendon disruption or to assist in reconstruction surgeries
- Part Number110045160
- Lot codes and dates
- UDI-DI10818674025772
- Lot Numbers
21E2821R1521R1621R3221R3522C1922E2922F1022F3522J0722J0822J0922J1822J1922J2722J2822J3322K2322K2422K3222K3322K3822M0322M0422M2422P0322P0922P1022P1122P1322P1422P16
- Amount recalled
- N/A
This recall covers 6 products
The agency filed one notice per product. You are reading one of them; the other 5 are below.
Where it was sold
- Nationwide
Why it was recalled
Drill guides might bend when axial forces are applied by surgeon during procedure and result in potential metal shavings due to friction with the drill bit.Risk to patients includes potential of injury.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new T.a.g. Medical Products Corporation, Ltd. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.