ADVIA Centaur Systems Insulin ReadyPack; Test Code: IRI; Catalog Number: 02230141, for in vitro diagnostic use in the…
Siemens Healthcare Diagnostics, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Siemens Healthcare Diagnostics, Inc · 333 Coney St · East Walpole, MA · 02032-1516
Is this the product you have?
- ADVIA Centaur Systems Insulin ReadyPack
- Test CodeIRI
- Catalog Number02230141, for in vitro diagnostic use in the quantitative determination of insulin in serum
- Lot codes and dates
- Kit Lots Ending
169170173174
- Kit Lots Ending
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
Current lots of ADVIA Centaur/XP/XPT/CP Insulin (IRI) assay recover approximately 40% higher than the World Health Organization (WHO) 1st IRP 66/304 standardization based on slope values. Other performance characteristics are being met per the assay Instructions for Use. Overall risk to health is negligible and Siemens is not recommending a review of previously generated results.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Siemens Healthcare Diagnostics, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.