Reported Jul 6, 2016 · Terminated

Ingenuity Core 128-Computed Tomography X-ray system Product Usage: The Ingenuity Core 128 scanner is a whole body…

Philips Medical Systems (Cleveland) Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Philips Medical Systems (Cleveland) Inc · 595 Miner Rd · Cleveland, OH · 44143-2131

Is this the product you have?

  • Ingenuity Core 128-Computed
  • Tomography X-ray system Product UsageThe Ingenuity Core 128 scanner is a whole body Computed Tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes
Lot codes and dates
  • Ingenuity Core128System Code #728323
  • Serial numbers
    • 320001
    • 320016
    • 320076
    • 320077
    • 320080
    • 320085
    • 320086
    • 320087
    • 320088
    • 320089
    • 320091
    • 320113
    • 336001
    • 336002
    • 336003
    • 336006
    • 336007
    • 336008
    • 336009
    • 336010
    • 336014
Amount recalled
19

This recall covers 10 products

The agency filed one notice per product. You are reading one of them; the other 9 are below.

Where it was sold

Why it was recalled

Philips Healthcare received reports from the field that certain Ingenuity Core128 systems running software version 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images. A patient rescan may be required if the images cannot be used for interpretation due to the swirl-like artifact.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Philips Medical Systems (Cleveland) Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jun 26, 2018 · Data as of Sep 19, 6:01 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.