Reported Jun 13, 2018 · Terminated
Y-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (003-C-01001); b) Large…
Orthofix, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Orthofix, Inc · 3451 Plano Pkwy · Lewisville, TX · 75056-9453
Is this the product you have?
- Y-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label
- Size (Part Number)
a) Small (003-C-01001)b) Large (003-C-02001)
- Lot codes and dates
(UDI)/Lot: a) (18056099647786) B1102163(b (18056099647793) B1106912
- Amount recalled
- 36
This recall covers 20 products
The agency filed one notice per product. You are reading one of them; the other 19 are below.
7 more are in the archive.
Where it was sold
Why it was recalled
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Orthofix, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.