Reported Aug 4, 2021 · Terminated

Heater Cooler Unit (HCU 30) The device supplies temperature-controlled water for cardioplegia heat exchangers and for…

Maquet Medical Systems USA

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Maquet Medical Systems USA · 45 Barbour Pond Dr · Wayne, NJ · 07470-2094

Is this the product you have?

  • Heater Cooler Unit (HCU 30) The device supplies temperature-controlled water for cardioplegia heat exchangers and for blood heat exchangers in extracorporeal circulation and for blankets with which patients can be warmed or cooled.
Lot codes and dates
  • HCU 30 Base Unit 100-120 V, Model Number 704629 I 70103.4371 HCU 30 Base Unit Canadian/US, Model Number 705501 I 70103.4653HCU 30 Base Unit 200-240 V, Model Number 939001 I 70102.8718 All serial numbers are affected.
Amount recalled
33

Where it was sold

Why it was recalled

There is the potential for contaminated water to enter parts of the Heater-Cooler Unit circuits. There is a potential risk of bacterial contamination including Mycobacteria species.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Maquet Medical Systems USA recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Aug 3, 2022 · Data as of Sep 19, 5:36 PM UTC