Reported Aug 20, 2025 · Ongoing

Baxter Novum IQ Large Volume Pump, Product Code REF 40700BAXUS

Baxter Healthcare Corporation

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Baxter Healthcare Corporation · 1 Baxter Pkwy · Deerfield, IL · 60015-4625

Is this the product you have?

  • Baxter Novum IQ Large Volume Pump, Product Code REF 40700BAXUS
Lot codes and dates
  • UDI/DI 05413765851797, All Serial Numbers
Amount recalled
36705 units

Where it was sold

  • Nationwide

Why it was recalled

Stop using this product now

Baxter is issuing an Urgent Medical Device Correction for the Novum IQ large volume pump (LVP) due to the potential for underinfusion when transitioning from a flow rate to a higher flow rate that is more than double. (e.g., rate change or bolus). The level of underinfusion is variable based on the current infusion rate, the duration the pump has been running at this flow rate, and the magnitude of the rate change. The longer the duration the pump has been running at the current infusion rate and the larger the magnitude of the rate change, the larger the underinfusion that would be experienced. Additionally, Baxter has identified an increase in customer reports of over and underinfusion potentially due to set misloading. Failure to properly load the tubing into the pump channel may result in the pump infusing at a rate higher or lower than programmed. Consistent with the instructions for use, customers should ensure that: 1) The door is fully open before loading the set. 2) The tubing is taut and loaded without slack in the pumping channel.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Baxter Healthcare Corporation recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 4:15 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.