Reported Jul 3, 2024 · Ongoing

Accu-Chek Guide (SC) Kit -Intended to quantitatively measure glucose in fresh capillary whole blood from the fingertip,…

Roche Diabetes Care, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Roche Diabetes Care, Inc. · 9115 Hague Rd · Indianapolis, IN · 46256-1025

Is this the product you have?

  • Accu-Chek Guide (SC) Kit -Intended to quantitatively measure glucose in fresh capillary whole blood from the fingertip, palm, and upper arm as an aid in monitoring the effectiveness of glucose control
  • Product REF Number08453071001
Lot codes and dates
  • GTIN00365702729100
  • Serial Numbers
    • US: 92339920445
    • 92340117408
    • 92339920116
    • 92339955415
    • 92340120006
    • 92339094787
    • 92340116052
    • 92340120057
    • 92339744998
  • Canada93040305146
Amount recalled
48 units

Where it was sold

  • Nationwide

Why it was recalled

The reason for recall is the meters may show an incorrect measurement unit (mmol/L rather than mg/dL) and may result in the wrong unit of measure. Which may subsequently cause an error in interpreting a result as approximately 18 times lower that the actual blood glucose. Potential immediate health consequences of the glucose meter providing a reading in the wrong units of measurement (UOM) may result in inappropriate rescue therapy for presumed hypoglycemia including urgent and repeat administration of carbohydrates. This in turn could lead to frequent episodes of hyperglycemia which is unrecognized and undertreated.Long range consequences of continued use of the affected meters could include mismanagement of the patient s diabetes for longer periods of time potentially resulting in a major health event (in general persistent and/or severe hyperglycemia) that may lead them to seek medical attention, and depending on the duration of hyperglycemia, could contribute to the development of microvascular (i.e., retinopathy, neuropathy, nephropathy) and/or macrovascular (i.e., myocardial infarction, ischemic stroke) complications of diabetes and could lead to fetal damage (abnormal fetal growth and altered organ development and maturation, fetal hypoxemia, spontaneous abortion, and congenital anomalies) in pregnant women. This issue was identified by three initial consumer complaints, with reports that the meters were presenting the incorrect unit of measure.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Roche Diabetes Care, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 4:52 PM UTC