Reported Sep 1, 2021 · Ongoing
Accu-Chek Aviva Plus Test Strip, Blood Glucose Monitoring System, Model numbers: 06908152001, 06908217001, 06908268001,…
Roche Diabetes Care, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Roche Diabetes Care, Inc. · 9115 Hague Rd · Indianapolis, IN · 46256-1025
Is this the product you have?
- Accu-Chek Aviva Plus Test Strip, Blood Glucose Monitoring System
- Model numbers
0690815200106908217001069082680010690834900106908373001
- Lot codes and dates
- Since this is a labeling correction, this recall is not lot specific and is applicable to all non-expired lots.
- Amount recalled
- 10,546,420 (US), 19,120,668 (OUS)
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
- Nationwide
Why it was recalled
Test strip vials may open while inside sealed cartons during shipment. An open vial might expose the test strips to humidity, which might damage the strips and result in inaccurate results, potentially leading to inappropriate therapy decisions and adverse health consequences.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Roche Diabetes Care, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.